NTEGRA platform: NTEGRA Aura: Specification
|
Scanning probe Microscopy
|
| STM/ AFM (contact + semi-contact + non-contact) / Lateral Force Microscopy / Phase Imaging/Force Modulation/ Adhesion Force Imaging/ Magnetic Force Microscopy/ Electrostatic Force Microscopy/ Scanning Capacitance Microscopy/ Kelvin Probe Microscopy/ Spreading Resistance Imaging/ Lithography: AFM (Force and Current), STM |
|
Specification
|
Scan type
|
| Scanning by sample |
Scanning by probe* |
| Sample size |
Up to 40 mm in diameter,
to 15 mm in height
|
Up to 100 mm in diameter,
up to 15 mm in height
|
| Sample weight |
Up to 100 g
|
Up to 300 g
|
| XY sample positiniong range |
5x5 mm
|
| Positioning resolution |
readable resolution-5 um
sensitivity-2 um
|
|
Scan range
|
100x100x10 um
3x3x2,6 um
|
100x100x10 um
50x50x5 um
|
|
Up to 150x150x15 um**(DualScanTMmode)
|
Non linearity, XY
(with closed loop sensors) |
0.1%
|
0.15%
|
Noise level, Z
(RMS in bandwidth 1000Hz) |
With sensors
|
0.04 nm (typically),
0.06 nm
|
0.06 nm (typically),
0.07 nm
|
|
Without sensors
|
0.03 nm
|
0.05 nm
|
Noise level, XY***
(RMS in bandwidth 200Hz) |
With sensors
|
0.2 nm (typically),
0.3 nm (XY 100 um)
|
0.1 nm (typically),
0.2 nm
|
|
Without sensors
|
0.02 nm (XY 100 um),
0.001 nm (XY 3 um)
|
0.01 nm
|
Closed-Loop
EquivalentTM |
Noise level, XY
(RMS in bandwidth 200Hz)
|
0.012 nm (XY 3 um)
|
|
Noise level, Z
(RMS in bandwidth 1000 Hz)
|
0.02 nm
|
|
Zoom accuracy
|
5% typically
|
| Optical viewing system |
Optical resolution
|
1 um
|
3 um
|
|
Field of view
|
4.5-0.4 mm
|
2.0-0.4 mm
|
|
Continuous zoom
|
available
|
available
|
|
Temperature control
|
Range
|
From RT to +150oC
|
|
Stability
|
±0.005°C (typically), ±0.01°C
|
| Vacuum system |
Pressure
|
10-2Torr
|
|
Vibration isolation
|
Active
|
0.7-1000 Hz
|
|
Passive
|
above 1 kHz
|
* Scanning head can be configured to serve as a stand-alone device for specimens of unlimited sizes.
** Optionally can be expanded to 200x200x20 мm.
*** Built-in capacitive sensors have extremely low noise and any area down to 50x50 nm can be scanned with closed-loop control.