Practice details

Practice details of Chemical Contrast Asquisition.

See practice details of topography acquisition, ZnSSe/GaAs laser structure.

Cantilevers CSC12 with rectangular beam and small force constant were used. When studying the friction forces on the heterostructure cleavage, the long side of the cantilever beam should be set parallel to the sample edge. To increase the sensitivity of mode, the signal of lateral deflection of the cantilever must be chosen close to zero. To avoid artifacts in LFM image due to coupling between vertical and lateral cantilever deflection, slow scanning speed and high feedback gain are preferable.

Specific parameters of the scan can be found in f5.opt and f5.mdt

On the next page description of the used lateral force mode basics are presented.